Annals of Clinical & Laboratory Science 31:205-210 (2001)
© 2001 Association of Clinical Scientists

Use of Microtiter Plates for Latex Agglutination Immunoassay of Serum Alpha2-Macroglobulin

Tetsuo Makino
Department of Clinical Chemistry, Kanagawa Prefectural College of Nursing and Medical Technology, Yokohama, Japan

Address correspondence to Tetsuo Makino, Ph.D., Department of Clinical Chemistry, Kanagawa Prefectural College of Nursing and Medical Technology, 1-5-1 Nakao, Asahi-ku, Yokohama 241-0815, Japan; tel 81 045 361 6141; fax 81 045 362 8785; e-mail dr_tetsu{at}xb3.so-net.ne.jp


    Abstract
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Acknowledgements
 References
 
A simple, sensitive turbidimetric immunoassay is described for analysis of alpha2-macroglobulin (AMG) in serum, using microtiter plates and a commercial kit that features a sensitive latex reagent. Using microwells as reaction vessels and the latex reagent made it possible to minimize the sample size and the amounts of reagents. The procedure involves two pipetting steps and a brief incubation (10 min) at room temperature; the assay measures AMG concentrations ranging from 0.12 to 5 g/L and is unaffected by icterus, hemolysis, lipemia, or rheumatoid factor, even at high concentrations. Recovery of AMG added to three pooled serum samples averaged 95–105%. CVs for replicate analyses ranged from 1.9 to 4.8% (within-run) and 6.1 to 9.3% (between-run). Based on paired analyses of 70 serum specimens, AMG concentrations obtained by this turbidimetric microtiter immunoassay were in close agreement (correlation coefficient = 0.99) with results obtained by an enzyme-linked immunosorbent assay (ELISA). Serum AMG concentrations in 48 healthy adults (mean ± 2SD) were 1.87 ± 1.00 g/L.

(received 4 December 2000; accepted 18 January 2001)

Keywords: Alpha2-macroglobulin, latex immunoassay, turbidimetry, microtiter plates

Abbreviations: AMG, alpha2-macroglobulin; SRID, single radial immunodiffusion; RIE, rocket immunoelectrophoresis; ELISA, enzyme-linked immunosorbent assay; TIA, turbidimetric immunoassay; PEG, polyethylene glycol; BSA, bovine serum albumin


    Introduction
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Acknowledgements
 References
 
Although measurements of alpha2-macroglobulin (AMG) concentrations in serum and urine have only limited clinical significance at present [1], accurate and sensitive methods are needed for research on the biological functions and pathophysiology of AMG. For example, measuring the partition of zinc between AMG and albumin in serum requires sensitive analyses of AMG concentrations in serum samples and zinc-binding protein fractions [2]. Several immunoassay techniques for AMG have been developed, including single radial immunodiffusion (SRID) [3,4], rocket immunoelectrophoresis (RIE) [5,6], radioimmunoassay [7,8], enzyme-linked immunosorbent assay (ELISA) [7,9,10], and turbidimetric immunoassay (TIA) using polyethylene glycol (PEG) [2,1114]. A commercial kit for urine AMG analysis has been introduced that involves a turbidimetric immunoassay, based on the agglutination of antibody-coated latex particles [15]. This technique has been used with an automated immunochemical analyzer [16]. The present paper describes an adaptation of the latex agglutination immunoassay, in conjunction with polystyrene microtiter plates and a microtiter plate reader, for the analysis of serum AMG concentrations. As shown previously with trace metal assays [17,18], this precise, convenient, and economical analytical approach provides significant reductions of sample volume and minimizes the required amounts of analytical reagents.


    Materials and Methods
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Acknowledgements
 References
 
Instrumentation.  The following instruments were employed: (1) a microtiter plate spectrophotometer with 670 nm interference filter and 10 nm half-bandwidth (model MPR-A4i, Tosoh Corp., Japan), (2) a micromixer for microtiter plates (model V-5, Thermal Chemical Industry, Japan), with mixing speed at setting 2.2, (3) microtiter plates with wells of 400 µl capacity (flat-bottom, 96-well type, #3881, Iwaki, Japan), (4) semiautomatic pipette of the positive displacement type for sampling (model 100, Microman-Bio, Gilson Instruments, USA), and (5) semiautomatic pipettes of the repeater type for dispensing reagents (model 4780, Eppendorf GmbH, Germany).

Reagents.  This serum assay uses a commercial kit for urine AMG analysis ("LX-Eiken-AMG," Eiken Chemical Corp., 5-26-20 Ouzi, Kita-ku, Tokyo 114-0002, Japan), which comprises three reagents: (1) "LX-buffer," containing HEPES (N-2-hydroxyethyl-piperazine-N’-2-ethansulfonic acid) buffer (0.05 mol/L) plus a surface-active agent; (2) "AMG-I" solution, containing bovine serum albumin (BSA); and (3) "AMG-II" suspension, containing rabbit anti-AMG IgG-coated latex particles. These reagents are stable at 4°C for one yr. Human Serum Protein Calibrator (X0908, Dako Corp., Shijo-dori, Sakyo-ku, Kyoto 600–8493, Japan), which contains AMG, 2.00 g/L, was used as stock standard solution. Working standards were prepared by diluting the stock standard to 100, 200, 400, 600, 1000 and 2000 ng/ml with LX-buffer.

Procedure.  Reagents, including water, were kept at room temperature during the AMG assay. Water (0.25 ml) was pipetted into the circumferential wells of the microtiter plate (ie, #1–12 of rows A and H, plus #1 and #12 of rows B to G), and into all unused wells. Serum samples were diluted 5000-fold with LX-buffer solution. Samples (40 µl) of reagent blank (water), working standards, and diluted serum samples were pipetted into respective wells, followed by addition of AMG-I solution (150 µl). The microtiter plate was then mixed for 30 sec. AMG-II solution (50 µl) was added and the plate was mixed for 45 sec. At 1 min (T1) and 10 min (T2) after mixing was begun, the absorbance of each well was measured at 670 nm. The increase in absorbance ({Delta}A = T2–T1) was directly proportional to the AMG concentration (ng/ml) in the working standards and diluted serums. Finally, the values for the diluted serums were multiplied by 0.005 to express the AMG concentration as g/L of undiluted serum.

Method comparisons.  Two comparison assays were used: a fully automated procedure using same reagents with an immunochemical analyzer (model LX-2200, Aloka, Japan) [15,16] and the ELISA procedure of Voller et al [9], with modifications [7].


    Results
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Acknowledgements
 References
 
Reaction time.  Fig. 1Go shows calibration curves for AMG immunoassays with reaction times that ended (T2) at 5, 10, 15, 20, 25, or 30 min after mixing. The reaction period that ended 10 min after mixing was chosen for routine use, since this calibration curve provided adequate sensitivity and was linear throughout the concentration range from 0 to 1000 ng AMG/ml.



View larger version (22K):
[in this window]
[in a new window]
 
Fig. 1. The effect of reaction time on the calibration curves for AMG immunoassays performed at 20°C. Absorbance measurements were made at T2 = 5 min (+), 10 min (solid circles), 15 min (x), 20 min (open triangles), 25 min (open squares), or 30 min (solid diamonds).

 
Reaction temperature.  Fig. 2Go shows calibration curves for AMG immunoassays that were performed at 17°C, 20°C, and 23°C. The slopes of the calibration curve were slightly different at the three reaction temperatures, but no significant differences were found in the AMG concentrations of 10 serum samples from patients, assayed at the three temperatures and calculated from the respective calibration curves. The observed AMG concentrations (mean ± SD) were: 17°C, 1.90 ± 0.87 g/L; 20°C, 1.88 ± 0.85 g/L; and 23°C, 1.90 ± 0.86 g/L.



View larger version (18K):
[in this window]
[in a new window]
 
Fig. 2. The effect of reaction temperature on the calibration curves for AMG immunoassays performed with T2 = 10 min. The incubation steps and absorbance measurements were at 17°C (open triangles), 20°C (solid circles), or 23°C (open squares).

 
Detection limit.  The analytical detection limit (corresponding to the reagent blank + 2SD) was 24.5 ng/ml (ie, 0.12 g AMG/L of serum). The linear range of the AMG assay was 25 to 1000 ng/ml (from 0.125 to 5 g/L of serum).

Interference effects.  Potential sources of interference were investigated by analyzing pooled serum samples that had been spiked with bilirubin, hemoglobin, triglycerides, or rheumatoid factor. Hemolysis (<=5 g/ L hemoglobin), icterus (<=0.5 g/L bilirubin as bilirubin ditaurate disodium salt, Porphyrin Products, USA) or lipemia (<=20 mmol/L triglycerides, as "Intralipos," Green Cross, Japan) did not affect AMG determinations. The serum AMG assay was unaffected by spiking with rheumatoid factor (<=500 IU/ml, as "High Level Check·RF," International Reagents, Japan).

Analytical recovery.  The analytical recovery of AMG was measured by adding AMG standard solutions (0.1 to 0.3 g/L) to three pooled serum samples that had been diluted 1:100 with LX-buffer. One volume of AMG standard solution was mixed with 9 volumes of each diluted serum. The recovery of added AMG averaged 95–105%. When 3 serum samples with initial AMG concentrations from 2 to 5 g/L were serially diluted, the results of AMG assays were linearily related to the calculated diluted concentrations.

Precision.  Data for within-run and day-to-day precision of serum AMG assays are listed in Table 1Go, based upon replicate analyses of two pooled serum samples with AMG concentrations of approximately 1.0 and 2.7 g/L. The respective coefficients of variation were 4.8 and 1.9% (within-run) and 9.3 and 6.1% (day-to-day).


View this table:
[in this window]
[in a new window]
 
Table 1. Precision of serum AMG determinations, based on replicate analyses of pooled samples of serum from patients.
 
Reference intervals.  Assays of AMG concentrations in serum samples from 48 healthy adults (both sexes, ages 20–54 yr) gave the following reference interval (mean ± 2SD): 1.87 ± 1.00 g/L.

Method comparison studies.  Analytical results for AMG concentrations in patients’ serums by the present method (Y-axis) are compared with those obtained with a fully automated immunoanalyzer, LX-2200 (n = 80, X-axis, Fig. 3Go), or by an ELISA assay (n = 70, X-axis, Fig. 4Go). The correlation coefficients (r) were 0.975 and 0.998, respectively (p <0.01).



View larger version (18K):
[in this window]
[in a new window]
 
Fig. 3. Correlation chart for measurements of AMG concentrations in 80 serum specimens from patients. Results obtained by the present method (Y-axis) are compared with those obtained with the same assay kit [15,16], using an automated analyzer (Aloka model "LX-2200")(X-axis). Pearson’s correlation coefficient (r) was 0.975 (p <0.01).

 


View larger version (17K):
[in this window]
[in a new window]
 
Fig. 4. Correlation chart for measurements of AMG concentrations in 70 serum specimens from patients. Results obtained by the present method (Y-axis) are compared with those obtained by an ELISA procedure [7,9] (X-axis). Pearson’s correlation coefficient (r) was 0.988 (p <0.01).

 

    Discussion
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Acknowledgements
 References
 
The various immunoassay techniques for AMG determination have their respective advantages and disadvantages. The SRID and RIE techniques give relatively poor precision and SRID requires a long reaction time. The ELISA method is sensitive and accurate, but is considerably more complex than other immunoassays. The kinetic TIA with PEG is simple, rapid, and precise, but shows nonspecific agglutination with IgM and interference from lipemic serums [19]. The TIA using latex particles is 10–100 times more sensitive than the TIA using PEG. However, the TIA using latex particles usually requires an expensive automated apparatus. Using latex agglutination reagents and microtiter plates, Collet-Cassart et al [20,21] assayed serum concentrations of placental lactogen and ß2-microglobulin. The present paper describes a TIA assay for serum AMG that uses a commercial kit for urine AMG in conjunction with microtiter equipment.

Collet-Cassart et al [20,21] employed end-point assays that require a longer running time, whereas the present method uses a fixed-time kinetic assay. This system is rapid and has a wide analytical range, but the time and the temperature during the latex agglutination reaction must be maintained accurately. Although differences in reaction temperature slightly changed the slope of the calibration curve, they did not affect the observed concentrations of serum AMG.

In TIA assays that use PEG, human serum albumin in the standards binds nonspecifically to polystyrene and glass tubes [22]. In TIA that involve latex agglutination, the antibody-coated latex particles are also adsorbed to the reaction vessels [23,24]. Moreover, in a previous study [25], serum proteins were adsorbed onto the surface of the polystyrene microwells, causing analytical interference. In the present method AMG-I solution, which contains BSA, is added to microwells before the addition of antibody-coated latex particles. This prevents the adsorption of latex particles to the microwell.

Icteric, hemolyzed, or lipemic serum samples did not affect the present assay for AMG, because the serum samples were prediluted by 1:5000 (v/v). The accuracy of the present serum AMG assay, evaluated by analytical recovery and dilution tests, appears satisfactory. The accuracy of AMG measurements by the present method is supported by the correlation coefficient of 0.99 with results by an ELISA assay.

The within-run precision (CV) of the present method (1.9–4.8%), is similar to that (1.9–2.9%) obtained by a kinetic TIA using PEG [2], and that (3.9–4.9%) obtained by laser nephelometry [14].

The author has recently employed the present technique, together with a similar assay for serum albumin [26], for measurements of serum AMG-bound zinc and albumin-bound zinc by affinity chromatography (paper submitted for publication).


    Acknowledgements
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Acknowledgements
 References
 
The author thanks the Eiken Chemical Corporation for the commercial kits for urine AMG analysis and and Tosoh Corporation for the microtiter plate reader.


    References
 Top
 Abstract
 Introduction
 Materials and Methods
 Results
 Discussion
 Acknowledgements
 References
 

  1. Silverman LM, Christenson RH. Amino acids and proteins. In: Tietz Textbook of Clinical Chemistry, 2nd ed (Burtis CA, Ashwood ER, Eds), Saunders, Philadelphia, 1994; pp 625–734.
  2. Foote JW, Delves HT. Distribution of zinc among human serum proteins determined by affinity chromatography and atomic-absorption spectrophotometry. Analyst 1983:108:492–504.[Medline]
  3. Mancini M, Carbonara AO, Heremans JF. Immunochemical quantitation of antigens by single radial immunodiffusion. Immunochemistry 1965;2:235–254.[Medline]
  4. Wilding P, Adham NF, Mehl JW, Haverback BJ. Alpha2-macroglobulin concentrations in human serum. Nature 1967;17:1226–1227.
  5. Laurell CB. Electroimmunoassay. Scan J Clin Lab Invest 1972;29:Suppl 124:21–37.
  6. Scott BJ, Burnett D. The effect of the protein content of diluents on peak height in rocket immunoelectrophoresis. Clin Chim Acta 1978;89:475–478.[Medline]
  7. Ishibashi H, Hayashida K, Okubo H. Alpha2-macroglobulin. In: Methods in Enzymology, vol 163 (Sabato GD, Ed), Academic Press, San Diego 1988; pp 485–495.
  8. Panrucker DE, Lorscheider FL. Radioimmunoassay of rat acute-phase alpha2-macroglobulin. Biochim Biophys Acta 1982;26:184–191.
  9. Voller A, Bidwell DE, Bartlett A. Enzyme immunoassays in diagnostic medicine: theory and practice. Bull World Health Org 1976;53:55–65.[Medline]
  10. Out TA, Jansen HM, Van Steenwijk RP, De Nooijer MJ, Van De Graaf EA, Zuijderhoudt FM. ELISA of ceruloplasmin and alpha2-macroglobulin in paired bronchoalveolar lavage fluid and serum samples. Clin Chim Acta 1987;165:277–288.[Medline]
  11. Whicher JT, Price CP, Spencer K. Immunonephelometric immunoturbidimetric assays for proteins. CRC Crit Rev Clin Lab Sci 1983;18:213–260.
  12. Spencer K, Coombes EJ, Wood PJ. The measurement of amniotic fluid alpha2-macroglobulin by fixed time kinetic immunoturbidimetry. J Clin Chem Clin Biochem 1983;21:133–137.[Medline]
  13. Schliep G, Felgenhauer K. Rapid determination of proteins in serum and cerebrospinal fluid by laser-nephelometry. J Clin Chem Clin Biochem 1978;16: 631–635.[Medline]
  14. Schmitz HU, Nachbar J, Asbeck F. The determination of antithrombin III, alpha2-macroglobulin, and alpha2-antiplasmin in plasma by laser nephelometry. J Clin Chem Clin Biochem 1980;18:221–225.[Medline]
  15. Kawashima K, Yamaguchi T, Sakabe K, Kubota N, Itoh Y. Development and evaluation of a latex agglutination reagent for determing urinary alpha2-macroglobulin [abstract]. Jap J Clin Lab Automat 1997;22:395.
  16. Ohisa N, Igarashi T, Kawahara S, Urushi T, Matsuki R, Suzuki H, Kawamura R, Yoshida K, Sasaki T. Human alpha2-macroglobulin: its measurement and clinical significance [abstract]. Jap J Clin Lab Automat 1998;23:342.
  17. Makino T. A sensitive, direct colorimetric assay of serum zinc using nitro-PAPS and microwell plates. Clin Chim Acta 1991;197:209–220.[Medline]
  18. Makino T. A colorimetric assay of trace metals in clinical samples using sensitive reagents and microwell plates. Bunseki 1998;49–52.
  19. Price CP, Newman DJ. Light-scattering immunoassay. In: Principles and Practice of Immunoassay, 2nd ed (Price CP, Newman DJ, Eds), Macmillan, London 1997; pp 443–480.
  20. Collet-Cassart D, Limet JN, Van Krieken L, De Hertogh R. Turbidimetric latex immunoassay of placental lactogen on microtiter plates. Clin Chem 1989;35:141–143.[Abstract/Free Full Text]
  21. Collet-Cassart D, Poncelet S, De Nayer P. Immunoturbidimetric assay of ß2 microglobulin using latex particles in microplates. J Immunol Methods 1991;142: 183–185.[Medline]
  22. Bakker AJ. Immunoturbidimetry of urinary albumin: prevention of adsorption of albumin; influence of other urinary constituents. Clin Chem 1988;34:82–86.[Abstract/Free Full Text]
  23. Hitachi Ltd., Instruments Business Promotion Dept., Medical Systems Section. The cleaning of individual reaction vessels in the Hitachi-7020 automated analyzer. Labo-Scope 1997;16:13.
  24. Ichihara F, Namikawa K, Hirohara T, Hiruta N. Basic study of CRP latex agglutination reagent with monoclonal antibody [abstract]. Jap J Clin Lab Automat 1999;24:445.
  25. Makino T, Kamada S, Fujisawa T. Microplate colorimetric assay for determination of zinc in 6.3 µl serum samples. Biomed Trace Elem Res 1999;10:265–266.
  26. Makino T, Sakabe K, Kamada S. Use of microtiter plates for determination of proteins in serum by latex agglutination immunoassay: albumin. J Anal Bio-Sci 2000; 23:457–461.




This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions