Annals of Clinical and Laboratory Science, Vol 25, Issue 3, 264-271
Copyright © 1995 by Association of Clinical Scientists
Inductively coupled plasma mass spectrometry for trace element analysis in the clinical laboratory
KL Nuttall,
WH Gordon,
and
KO Ash
Inductively coupled plasma mass spectrometry (ICP-MS) is a relatively new technique for trace element analysis. The basic operating principles of ICP-MS are described and our experience with this technique in a clinical setting is discussed for the analysis of serum, whole blood, and urine. Advantages to ICP-MS include the favorable detection limits (0.01 to 0.1 micrograms/L for many elements), simple specimen preparation, high throughput (about 40 specimens per hour), and the ability to measure more than one element simultaneously. A major disadvantage is the high capital cost of the instrumentation. Heavier elements, such as lead, are well-suited for ICP-MS analysis, whereas lighter elements are prone to more interferences. Lighter elements which are not amenable to assay by ICP-MS include chromium and iron. The ability to measure isotopes is a major advantage for mass spectrometry methods and has the potential to expand the usefulness of trace element analysis.